ELECTRONIC DILATOMETERS SERIES L76

(new version, also with differential thermal analysis, DTA)

• Interchangeable thermocouples and measuring systems.
• Sinter process evaluation.
• Glass transition ad softening point evaluation.
• Softening point determination and system shut down.
• Linear thermal expansion evaluation.
• Several system correction features.
• Automatic zero point adjustment.
•Auto scheduler for up to 16 uninterrupted runs.

Software Description

• Possibility to record a zero curve for correcting the measuring system rod and specimen
container expansion (quartz, alumina, graphite, etc. ).
• Storing the value expansion of the rod material.
• The program makes it possible to record the differential, the differential with reference, or the absolute measurement of one or two dilatoreters.
• During measurement, the curve are displayed on the video with the indication of expansion and temperature.
• During the processing, can be printed the curves: expansion/derivatived of expansion as a temperature function;temperature and expansion as a time function; coefficent of expansion as temperature function.
• It is possible to program windows and print values only within pre-set limits.
• At pre-set intervals on the printer can be printed: the temperature/ the expansion/the expansion coefficent/ the percentage of expansion over the initial length of the sample, also in the table form.
• The software, furthermore, is suitable to make a DTA simulation test, through matematical calculation of the klin adjustment.

Laboratories